Introduction
A gas detector that has not been recently checked is a gadget, not a safety instrument. Sensors drift, age and can be poisoned; the only way to know an alarm will actually fire is to expose the instrument to a known gas. Calibration and bump testing are therefore the back office of gas safety — invisible until something goes wrong, and essential.
This guide explains the difference between a bump test and a full calibration, how often each should be done, the gases and equipment involved, and how to keep defensible records for audits.
What Is a Bump Test?
A bump test briefly applies a known concentration of test gas to the detector to confirm that the sensor responds and the alarms (audible, visual, vibration) actually activate. It does not adjust the reading. It answers a simple question: does this instrument still react at all? Best practice, widely recommended by safety-equipment associations and manufacturers, is to perform a bump test before each day’s use for portable instruments.
What Is a Calibration?
A calibration (or span check/adjustment) exposes the instrument to a known gas concentration and trims the electronics so the displayed reading matches the true value. A zero calibration in clean or zero air establishes the baseline; a span calibration sets the response to the target gas. A calibration corrects drift and confirms accuracy, not just response.
How Often Is Enough?
The widely accepted baseline — reflected in manufacturer manuals and safety-association guidance — is: bump test before each day’s use, and full calibration at least every six months. Shorten the interval — or recalibrate immediately — when:
- A bump test fails or the response is weak.
- The detector is dropped, submerged, or exposed to a very high concentration or a known poison.
- The sensor cell is replaced.
- The environment is exceptionally dusty, humid, corrosive or temperature-extreme, which accelerates drift.
Fixed detectors follow the same principle on a longer, scheduled cycle — typically a periodic bump/function check and a six-month (or manufacturer-specified) calibration, tightened for harsh outdoor service. Always follow the instrument manual and the site’s approved maintenance procedure.
Bump Test vs Calibration
| Bump test | Calibration | |
| Purpose | Confirm response & alarms work | Adjust reading to true value |
| Frequency | Before each day’s use (portable) | At least every 6 months, or after failure |
| Adjustment | None | Trims zero / span |
| Cal gas needed | Yes, short exposure | Yes, controlled flow, certified cylinder |
| Time per unit | Minutes | Longer, documented |
A Typical Calibration Procedure
A routine calibration follows the same sequence every time, which keeps results repeatable. First, place the instrument in fresh, breathable air (or certified zero air) and perform the zero adjustment so the baseline reads zero for toxic channels and 20.9 % for oxygen. Second, apply the certified span gas at the manufacturer-specified flow rate through a demand-flow or fixed-flow adapter and wait for the reading to stabilise. Third, adjust the span so the display matches the known cylinder concentration. Finally, document the result and confirm the alarm setpoints still trip at the expected concentrations. Cutting the wait-for-stabilise step is a common cause of drift between calibrations.
Calibration Gas, Cylinders and Records
Use certified, traceable calibration gases matched to the detector’s target gases and ranges — and within shelf life. A regulator and fixed-flow demand flow fitting ensure the sensor actually sees the gas. Keep records of every bump and calibration: date, unit ID, gas used, result, and the technician. These records are what an auditor or regulator asks for first, and they are the evidence that the safety programme is real.
For a fleet of portables, organise the work on a schedule rather than reacting to failures: tag each unit with its last calibration date, rotate units through a calibration bench, and retire units that repeatedly fail. For fixed detectors, combine the periodic function check with the planned maintenance route so the visit is not doubled up. Either way, the discipline of bump-before-use plus scheduled calibration is what keeps a safety-critical instrument trustworthy.
One further point: calibration itself can be dangerous if done carelessly. Cylinders are pressurised, calibration gas includes toxic components such as H2S, and hydrogen and fuel mixtures are flammable. Store and handle cylinders upright, secured, away from heat, and vent exhaust to a safe, well-ventilated area — never into a confined calibration booth without adequate ventilation.
If a detector cannot be calibrated back to spec, that is information, not an inconvenience. It usually means the sensor is exhausted, contaminated or the instrument has suffered damage. Do not paper over the problem by resetting alarms or ignoring the failed adjustment; take the unit out of service and replace the affected sensor, then verify with a fresh bump test before returning it to the field.
Practical Tools That Reduce Downtime
Field service is easier when the instrument supports it. The Hangwei HD9300 fixed detector offers IR remote calibration, so technicians can calibrate without opening the hazardous enclosure or fighting mounting height. The HD620 portable clearly indicates alarm state and uses common four-gas calibration gas mixtures.
Fixed transmitter: HD9300 Fixed Gas Detector (IR remote calibration)
Portable: HD620 Four-In-One Pump Detector
FAQ
What is the difference between a bump test and a calibration?
A bump test confirms the sensor and alarms respond when exposed to gas. A calibration actually adjusts the reading against a known concentration to correct drift.
How often should I bump test my gas detector?
Best practice for portable detectors is a bump test before each day’s use. Fixed detectors should be function/bump checked on a scheduled basis.
How often should I calibrate?
At least every six months as a baseline, and immediately after a failed bump, a drop, a high-concentration exposure, sensor replacement, or in particularly harsh service.
What happens if a bump test fails?
Take the instrument out of service. Perform a calibration; if it still cannot respond, replace the sensor cell or the instrument, before allowing it back into use.
Do I need special gas cylinders?
Yes. Use certified, traceable calibration gases matched to the detector’s gases and ranges, within their shelf life, with a suitable regulator and demand-flow fitting.
Get Expert Guidance
Hangwei Sensors designs detectors with maintainability in mind — IR remote calibration on the fixed HD9300 and clear alarm indication on the HD620 portable — and can advise on calibration planning, spare sensors and global distributor support. With 20+ years of manufacturing, 20 patents and 10 production lines, we serve Russia, Central Asia and the Middle East.
Fixed detector: HD9300 Fixed Gas Detector
Portable: HD620 Four-In-One Pump Detector
Spare sensing elements: Modular Sensor
Email: hw@hangweisensors.com · Phone: +86 371 6667 0886 · Website: https://www.hangweisensors.com/
Sources & References
[1] ISA — maintenance & calibration of instrumentation
[2] CDC NIOSH — respiratory / gas monitoring guidance
[3] IEC — IEC 60079 / IECEx series
[4] Hangwei Sensors — HD9300 Fixed Detector